Table 1 Average and expanded uncertainty of the detection efficiency reported by each participant laboratory for the mean photon per pulse between 0.01 and 2.4 (1100 photon/s and 193,000 photon/s)

From: A study to develop a robust method for measuring the detection efficiency of free-running InGaAs/InP single-photon detectors

Laboratory Participant detection efficiency, \(\eta _{p}\) Expanded Uncertainty, (rel.)
CMI 0.1043 0.0056 (5.3%)
INRIM 0.1052 0.0050 (4.8%)
NPL 0.1090 0.0054 (5.0%)
PTB 0.1075 0.0024 (2.7%)
\