Skip to main content

Table A2 Uncertainty budget of the SPAD detection efficiency measurement performed by INRIM for a mean photon number \(\mu =0.146\)

From: A study to develop a robust method for measuring the detection efficiency of free-running InGaAs/InP single-photon detectors

Measurand

Value

Standard Uncertainty

Type of distribution

Sensitivity coefficient

Contribution

Contribution (%)

Planck’s constant, h, [J s]

6.6260701510−34 J s

0

Rectangular

\( 1.603\cdot 10^{32} \)

0

0

Wavelength, λ, [m]

1.550510−6 m

610−11 m

Rectangular

\( -6.85\cdot 10^{4} \)

−4.1110−6

3.3610−4

Speed of light, c, [m/s]

299.79245810−6 m s−1

0

Standard

3.5410−10

0

0

InGaAs SPAD Count rate, \(\rho _{\mathit{click}}\), [s−1]

2542 s−1

8 s−1

Student

6.6610−5

5.3310−4

5.64

InGaAs SPAD Dark Count rate, \(\rho _{\mathit{Dark},\mathit{click}}\), [s−1]

910 s−1

3 s−1

Student

−6.4210−5

−1.9310−4

0.74

Power meter measurement, \(P_{0}\), [W]

1.1510−8 W

3.0010−11 W

Standard

\( -9.25\cdot 10^{6} \)

−2.7810−4

1.53

Attenuator factor, α

1.7110−7

8.4210−10

Standard

\( -6.21\cdot 10^{5} \)

−5.2310−4

5.44

Connect/disconnect fibre factor, \(F_{\mathrm{fconn}}\)

1

0.005

Rectangular

0.106

5.3110−4

5.61

InGaAs SPAD Dead time, D, [s]

10.22010−6 s

5.9010−9 s

Rectangular

252.68

1.4910−6

4.4210−5

Laser frequency, \(f_{\mathit{laser}}\), [s−1]

110103 s−1

63.51 s−1

Rectangular

−1.8410−8

−1.1710−6

2.7210−5

Power meter calibration factor, \(F_{\mathrm{cal}}\)

1.046

0.020

Standard

−0.102

−2.0210−3

81.04

Detection efficiency, η

0.1063

0.0022

Standard